Troubleshooting
If an error message appears on the screen during normal operation, print the error and fax to
Polar/HealthFirst Corporation Technical Support at 505/344-1200. To print a fax technical
request form from TriFIT, go to Help and Technical Support. Complete the form and click
print. You can then fax this form for quicker response to your question.
On-Line Peripherals
PLATFORM
Platform is giving low or high readings
1. Re-calibrate (Defaults/Calibration) being sure to follow the instructions on the
screen exactly.
2. Use a person who has a known weight between 150 and 190lbs.
3. Make sure the person DOES NOT step on the scale before it is zeroed. This will
cause the platform to be mis-calibrated.
4. Make sure the person stands very still. Any movement from side to side will cause a
mis-calibration.
5. Go to a test and perform an actual body weight with the same person that was used
to calibrate the scale. The weight should be within two pounds.
6. If the weight is not within two pounds, re-calibrate and re-measure the body weight.
7. If the weight is still varying by more than two pounds, call the Polar/HealthFirst
Corporation Technical Support Line (505-344-9679).
8. Write down any error codes, as this will be helpful in correcting the problem.
CALIPERS
On-Line Calipers are giving inaccurate readings
1. Re-calibrate as instructed on the screen. (Defaults/Calibration).
2. Go into a 3-site Skinfold test.
3. Zero the calipers.
4. Place the 25.4mm calibration dowel in the jaws of the calipers lengthwise.
5. Press the “Measure” switch on the calipers.
6. The reading on the screen should be within 24.5mm and 26.0mm. Do this 3 times
and note each reading.
7. If the readings are not within 24.5mm and 26.0mm, re-calibrate and repeat.
8. If the dial on the calipers do not read exactly 25.4mm when the dowel is in the
calipers, do not worry! The dial on the caliper is not as accurate as the digital
reading being sent to the computer.
9. If still not within the above values, call the Polar/HealthFirst Corporation Technical
Support Line (505-344-9679).